- Products Search > Life Science Lab Instruments > MCE-202(MultiNA) > (MCE) There is a problem with Size Accuracy
(MCE) There is a problem with Size Accuracy
(MCE) There is a problem with Size Accuracy
- Category :
Answer
« Analysis performance deviates from specification around the lower end of the size range. »
• The ladder size and target fragment size are too far separated from one another.
→Use an optional ladder or user ladder instead.
Adjust the diluted ladder solution according to the salt concentration of the sample.
Adjust the diluted ladder solution according to the salt concentration of the sample.
« The estimated size value is different from the actual size. »
• The sample concentration is too high.
→A high sample concentration can cause a small estimated size value. Dilute the sample.
• The nucleic acid in the sample was formed into structures.
→If the nucleic acid in the sample is not linear, the estimated size value may differ from the actual size.