FAQ - Frequently Asked Questions
- Products Search > Life Science Lab Instruments > MCE-202(MultiNA) > (MCE) There is a problem with peak recognition
(MCE) There is a problem with peak recognition
(MCE) There is a problem with peak recognition
- Category :
Answer
« Peaks other than the marker peaks are recognized as marker peaks. »
• Another peak near the marker is mistakenly recognized as the marker.
→Set the correct marker peak using the [Manual Edit Mode] accessible from the [Reanalysis] menu in the Viewer window.
« The ladder fragment peaks are confused with inappropriate peaks. »
• A peak not derived from the ladder is mistakenly recognized as a ladderderived peak.
→Set the correct fragment peak using the [Manual Edit Mode] accessible from the [Reanalysis] menu in the Viewer window.
* If the analysis performance check or the rudder monitor fails the judgment result, replace the microchip with a new one or perform chip cleaning.
Click here for "(MCE) Chip Cleaning".
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.