• Font Size
  • S
  • M
  • L
  • No : 8365
  • Date of Issue : 2022/02/17 15:39
  • Print

(MCE) Peak does not detect

(MCE) Peak does not detect
Category : 

Answer

« None of the microchips detect peaks for markers or samples. »
• Incorrect usage of microchips, reagent, rinse water, glass rinse water bottle, or waste container.
→Verify that the microchips, reagent, rinse water, glass rinse water bottle, and waste containers have been used properly. If the problem reappears after confirmation, contact a Shimadzu sales or service representative.
 
• The sample salt concentration is too high.
→Check that the salt concentration in the sample is within the permissible salt concentration range of the reagent kit.
 
• The microchip surface is contaminated.
→Check the microchip surfaces for contamination. If they are contaminated, rinse them.
  Click here for "(MCE) Inspecting and Washing Microchip Reservoirs (manually)".
 
« Upper marker or lower marker peaks are detected, but the sample/ladder peaks are not. »
• The sample concentration is lower than the concentration limit of detection.
→Check the specifications for the minimum concentration for the reagent kit used.
 
• The sample size is outside of the size range for the separation buffer.
→Check the size range specifications for the reagent kit used.
 
• (On-chip mixing mode)The volume of sample solution in the sample tubes is less than the prescribed quantity.
→For on-chip mixing, provide at least 5μL of sample. Attach the aluminum seal to the sample plate to prevent evaporation.
 
• (On-chip mixing mode)The sample solution evaporates inside the instrument, reducing the volume.
→Attach the aluminum seal to the sample plate.
 
« Only lower marker peaks are detected. Upper marker and sample/ladder peaks are not detected. »
• Dye is not added to the separation buffer. The dye added to the separation buffer has deteriorated.
→Use new separation buffer to which freshly prepared dye solution is added.
 
« Although the lower marker and sample/ladder peaks are detected, the upper marker is not detected. »
• The marker solution has degraded.
→Check whether the solution was stored at the specified temperature, whether the solution was excessively agitated in a vortex mixer, etc., or whether the solution is still within the recommended usage period.
 
• The sample concentration or salt concentration in the sample is too high.
→Dilute the sample. Desalt the sample.
 
• Enzymes in the sample were not inactivated.
→Inactivate the enzyme activity according to the manufacturer's instruction.
 
« The sample/ladder peaks are detected, but the marker peaks are not. »
• Premix mode
→The marker solution has not been mixed with the sample/ladder. Verify that the marker solution was mixed with the sample/ladder.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.