(MCE) Replacement and Maintenance of Suction Filter
(MCE) Replacement and Maintenance of Suction Filter
- Category :
Answer
« Guidelines for Procedure »
• When a significant number of bubbles are present in the rinse water suction line.
• When the suction filter appears to be dirty.
• When a significant number of bubbles are present in the rinse water suction line.
• When the suction filter appears to be dirty.
« Requirements »
• Ultrasonic cleaner
• 2-propanol
• Beaker
• ELEMENT, SUS FILTER: Part No. 228-45707-91 (if replacement is required)
• 2-propanol
• Beaker
• ELEMENT, SUS FILTER: Part No. 228-45707-91 (if replacement is required)
« Procedure »
1. Remove the suction filter from the glass rinse water bottle, and pull the suction filter off of the suction tubing.
2. Place the suction filter in a beaker filled with 2-propanol, and clean it for 5 minutes using an ultrasonic cleaner.
2. Place the suction filter in a beaker filled with 2-propanol, and clean it for 5 minutes using an ultrasonic cleaner.
3. Replace the suction filter on the suction tube, and return it to the glass rinse water bottle.
CAUTION: Insert the suction filter into the suction tube marked with an "F".
4. Select [Wash] on the [Instrument] menu in the MultiNA window and click [Probe Rinse].
A probe rinse is executed.
A probe rinse is executed.
5. Verify that no bubbles accumulate in the suction tubing.
If bubbles are still accumulating in the suction tubing, replace the suction filter with a new one (ELEMENT, SUS FILTER).
If bubbles are still accumulating in the suction tubing, replace the suction filter with a new one (ELEMENT, SUS FILTER).