(MCE) Replacement and Inspection of Pump Cartridge
(MCE) Replacement and Inspection of Pump Cartridge
NOTE: Pump cartridges are not compatible between different pumps. Check the type of pump cartridge installed on the instrument and replace it with an identical one.
Part Number: 042-00405-13 CASSETTE HEAD
Part Number: 292-96682 CASSETTE HEAD
« Purpose »
Extended use wears out the pump cartridge rotor and may cause a reduction in flow rate or tube failure.
Replace the pump cartridges when the guidelines (standard interval) for replacement have been reached.
Replace the pump cartridges when the guidelines (standard interval) for replacement have been reached.
« Replacement Guidelines »
When the pump cartridge operation time displayed in the [Parts Maintenance] window exceeds the guidelines (standard interval) for replacement.
It is also necessary to replace the pump cartridge when there is at least one "fail" in the analysis performance inspection.
It is also necessary to replace the pump cartridge when there is at least one "fail" in the analysis performance inspection.