FAQ - Frequently Asked Questions
- Products Search > Life Science Lab Instruments > MCE-202(MultiNA) > (MCE) Replacement and Inspection of Syringe
- No : 8345
- Date of Issue : 2022/02/16 08:42
(MCE) Replacement and Inspection of Syringe
(MCE) Replacement and Inspection of Syringe
Part Number: 228-25237-04 Syringe
« Purpose »
If the instrument is used for an extended period of time, air tightness is reduced by syringe barrel wear. If air tightness is not improved by replacing the plunger, then replace the syringe.
If the instrument is used for an extended period of time, air tightness is reduced by syringe barrel wear. If air tightness is not improved by replacing the plunger, then replace the syringe.
« Replacement Guidelines »
Perform this procedure in either of the following situations:
• When either the number of analyses or replacement date for the syringe, indicated in the [Parts Maintenance] window, exceeds the guidelines (standard interval) for replacement.
• If leakage is discovered below the plunger seal.
(The syringe cover must be removed if the syringe is found to be leaking.)
• When either the number of analyses or replacement date for the syringe, indicated in the [Parts Maintenance] window, exceeds the guidelines (standard interval) for replacement.
• If leakage is discovered below the plunger seal.
(The syringe cover must be removed if the syringe is found to be leaking.)