FAQ - Frequently Asked Questions
- Products Search > Life Science Lab Instruments > MCE-202(MultiNA) > (MCE) Replacement and Inspection of Plunger
- No : 8344
- Date of Issue : 2022/02/16 08:42
(MCE) Replacement and Inspection of Plunger
(MCE) Replacement and Inspection of Plunger
The syringe consists of a plunger and syringe barrel (glass cylinder). A PTFE plunger seal is attached to the plunger tip to improve air tightness.
Syringe (Left), Syringe Barrel (Center) and Plunger (Right). The Tip of the PTFE Plunger Seal is Circled
Part Number: 228-25237-14 Plunger ASSY,1750
« Purpose »
If the instrument is used for an extended period of time, the air tightness of the plunger seal is reduced by wear. Replace the plunger to restore the air tightness.
NOTE: It is not possible to replace only the plunger seal.
If the instrument is used for an extended period of time, the air tightness of the plunger seal is reduced by wear. Replace the plunger to restore the air tightness.
NOTE: It is not possible to replace only the plunger seal.
« Replacement Guidelines »
Perform this procedure in either of the following situations:
• When either the number of analyses or replacement date for the plunger, indicated in the [Parts Maintenance] window, exceeds the guidelines (standard interval) for replacement.
• If leakage is discovered below the plunger seal.
(The syringe cover must be removed if the syringe is found to be leaking.)
• When either the number of analyses or replacement date for the plunger, indicated in the [Parts Maintenance] window, exceeds the guidelines (standard interval) for replacement.
• If leakage is discovered below the plunger seal.
(The syringe cover must be removed if the syringe is found to be leaking.)
Please select a maintenance item.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.