(MCE) Chip Cleaning
(MCE) Chip Cleaning
« Purpose »
To remove components adhered in the microchip's channels using chip cleaning solution.
« Guidelines for Procedure »
• When the UM theoretical plate number or UM time fail in the analysis performance inspection• When the separation performance is determined to deteriorate as the upper marker (UM) peak widens
« Requirements »
• Chip cleaning kit (Part Number: 292-35925-91)
• Buffer bottle
• Open-hole screw cap
• Micropipette and pipette tip
• Milli-Q ultrapure water
• Wipes
• Buffer bottle
• Open-hole screw cap
• Micropipette and pipette tip
• Milli-Q ultrapure water
• Wipes
WARNING: Always read the instruction manual provided with the chip cleaning kit before handling the contents of the kit.
The instruction manual for the chip cleaning kit describes safety precautions that are important to ensure safe use.
The instruction manual for the chip cleaning kit describes safety precautions that are important to ensure safe use.