FAQ - Frequently Asked Questions
- Products Search > Life Science Lab Instruments > MCE-202(MultiNA) > (MCE) Maintenance of Sample (Extra Sample) Stand
(MCE) Maintenance of Sample (Extra Sample) Stand
(MCE) Maintenance of Sample (Extra Sample) Stand
- Category :
Answer
« Guidelines for Procedure »
When adhered substances are present on the sample stand or extra sample stand.
« Requirements »
• Wipes
• Milli-Q ultrapure water
• Milli-Q ultrapure water
« Cleaning Procedure »
1. Perform the procedure in "Moving All Axes to Home Position".
Click here for "(MCE) Moving All Axes to Home Position".
2. Wipe off any adhered substances with a wipe soaked in Milli-Q ultrapure water.
3. Then remove any residual water with a dry wipe.
3. Then remove any residual water with a dry wipe.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.