- Products Search > Life Science Lab Instruments > MCE-202(MultiNA) > (MCE) Maintenance of Air Cylinder Seal
(MCE) Maintenance of Air Cylinder Seal
(MCE) Maintenance of Air Cylinder Seal
- Category :
Answer
« Purpose »
To remove any substances adhered to the air cylinder seal.
Salt or other substances adhered to the air cylinder seal will reduce the efficiency of the separation buffer transfer process.
Salt or other substances adhered to the air cylinder seal will reduce the efficiency of the separation buffer transfer process.
« Guidelines for Procedure »
• When adhered substances are present on the air cylinder seal.
• When a seal mark is evident on a microchip surface after analysis is complete.
• When a seal mark is evident on a microchip surface after analysis is complete.
« Requirements »
• Wipes
• Milli-Q ultrapure water
• Milli-Q ultrapure water
« Cleaning Procedure »
1. Perform the procedure in "Moving All Axes to Home Position".
Click here for "(MCE) Moving All Axes to Home Position".
2. Wipe the tip of the air cylinder seal with a wipe soaked in water.
CAUTION:
• Do NOT scratch the air cylinder seal.
• Do NOT apply unnecessary force to the tip of the aspirator.
• Do NOT apply unnecessary force to the tip of the aspirator.
3. Then remove any residual water with a dry wipe.