(MCE) Maintenance of Air Cylinder Seal

(MCE) Maintenance of Air Cylinder Seal
Category : 

Answer

« Purpose »
  To remove any substances adhered to the air cylinder seal.
  Salt or other substances adhered to the air cylinder seal will reduce the efficiency of the separation buffer transfer process.
 
« Guidelines for Procedure »
  • When adhered substances are present on the air cylinder seal.
  • When a seal mark is evident on a microchip surface after analysis is complete.
 
« Requirements »
  • Wipes
  • Milli-Q ultrapure water
 
« Cleaning Procedure »
1. Perform the procedure in "Moving All Axes to Home Position".
  Click here for "(MCE) Moving All Axes to Home Position".
2. Wipe the tip of the air cylinder seal with a wipe soaked in water.
  CAUTION:
    • Do NOT scratch the air cylinder seal.
    • Do NOT apply unnecessary force to the tip of the aspirator.
3. Then remove any residual water with a dry wipe.
  