(EZ-Test) Calibrating Test Force (E-CAL)
(EZ-Test) Calibrating Test Force (E-CAL)
- Category :
Answer
Electronically calibrates (E-CAL) the test force. Calibrating the test force with E-CAL uses precision resistors built into the calibration cable to apply a simulated load on the load cell and calibrate the test force amplifier
[[Setting Procedure]]

[Setting Value]
Force: Electronically calibrates (E CAL) the test force.
CAUTION: Calibrate test force by E-CAL with the load cell in the unloaded state.
Wait at least 15 minutes after switching the load cell power ON, before calibrating test force by E-CAL.
Wait at least 15 minutes after switching the load cell power ON, before calibrating test force by E-CAL.
NOTE: E-CAL takes about 30 seconds. Please wait without operation.