(EHF) The set maximum test force cannot be loaded / The set displacement amplitude does not appear
(EHF) The set maximum test force cannot be loaded / The set displacement amplitude does not appear
Energy Saving Power Unit Software Trouble
Windows Software for 4830
Answer
The maximum testable condition is a reference value. We do not guarantee the actual test. Depending on the state of the device, the value may not be. If the displacement amplitude is calculated to be less than 1 mm, --- is displayed.
Maximum test condition: The maximum value in the automatic calculation of test force is the dynamic test force value, and the maximum value when input is the static test force value. If you press the buttons for displacement amplitude and maximum test force alternately, the maximum test conditions may change in order due to the effect of rounding of numerical values.
Maximum test condition: The maximum value in the automatic calculation of test force is the dynamic test force value, and the maximum value when input is the static test force value. If you press the buttons for displacement amplitude and maximum test force alternately, the maximum test conditions may change in order due to the effect of rounding of numerical values.