(EHF) A test cannot be performed

(EHF) A test cannot be performed
Before/after testing Software Trouble
Windows Software for 4830
Category : 

Answer

• When usi ng multiple controllers, the set number of synchronously con nected 4830 controllers must be the same as the number of controllers connected by synchronous cable. For example, if there are three controllers and only two are used, connect only the two contro llers by synchronous cable and set the number of synchronous connections to two.
  If the number of actually connected controllers is less than the set number, the system will wait for synchronization of the controller that is not connected at the beginning of the test and thus it will not be possible to execute test ing.
• When using an optional test function, multiple tests cannot be executed simultaneously.