(UH) Raising / Lowering the Table
(UH) Raising / Lowering the Table
UH-X/Xh Series, UH-FX/Xh Series, X-Type Mesurement Controller
Answer
Set the Manual mode before manually raising/lowering the table.
1. Press the control mode selector (MANU.) key to set the Manual mode.
2. Install the upper and lower platens.
3. Lower the lower crosshead until the distance between the upper and lower platens becomes approx. 1 to 2 mm.
4. Turn the load control knob gradually toward “OPEN” to raise the ram. Stop raising the ram when the upper and lower platens make contact.
2. Install the upper and lower platens.
3. Lower the lower crosshead until the distance between the upper and lower platens becomes approx. 1 to 2 mm.
4. Turn the load control knob gradually toward “OPEN” to raise the ram. Stop raising the ram when the upper and lower platens make contact.
NOTE: Allowing the upper and lower platens to make contact suddenly gives a strong impact on the testing machine. When the platens are near, be sure to slowly operate the load control knob.
5. A test force is applied to the testing machine when the upper and lower platens are in contact. Increase the test force to close to the rated test force by turning the load control knob gradually toward “OPEN”. At this time, turn the load control knob to the position where the test-force value becomes constant or falls slightly.
6. Press the control mode selector (MANU.) key to set the Auto mode.
7. Although the test force starts falling, do not operate the load control knob at this point.
8. Touch the [System] menu key on the LCD display. The System screen appears.
6. Press the control mode selector (MANU.) key to set the Auto mode.
7. Although the test force starts falling, do not operate the load control knob at this point.
8. Touch the [System] menu key on the LCD display. The System screen appears.
9. Touch the [Maintenance] key.
10. To correct the HOLD position of the load control knob, select the [Special Set] tab, and touch the [Adjust] key.
11. To initialize the corrected value, press the [Reset] key. This operation allows you to initialize the HOLD position of the load control knob.