- Products Search > Materials Testing & Inspection > Hardness Tester > Micro Hardness Tester(HMV) > (HMV) Replacing the Illumination Lamp
(HMV) Replacing the Illumination Lamp
(HMV) Replacing the Illumination Lamp
HMV-G30, HMV-G20
HMV-G31, HMV-G21
HMV-G31-FA, HMV-G-FA
HMV-G31-XY, HMV-G-XY
HMV-G31-XY, HMV-G-XY
Answer
Part Number: 347-24149-02 LED LAMP ASSY
1. Remove the top cover fastening screws, and remove the top cover.
2. Remove the fastening screws. Remove the lamp assembly, and then attach a new lamp assembly.
NOTE: During replacement, be careful that the lamp assembly cable does not get caught on the lever.
2. Remove the fastening screws. Remove the lamp assembly, and then attach a new lamp assembly.
NOTE: During replacement, be careful that the lamp assembly cable does not get caught on the lever.
3. Attach the top cover, and turn ON the instrument.
4. Check the field of view. If the entire field of view is bright, fasten the top cover with the screws. If the field of view is dark, loosen the fastening screws, and move the lamp assembly to adjust the position.
4. Check the field of view. If the entire field of view is bright, fasten the top cover with the screws. If the field of view is dark, loosen the fastening screws, and move the lamp assembly to adjust the position.