(LC) Cleaning the extraction vessel (SFE)

(LC) Cleaning the extraction vessel (SFE)
Category : 

Answer

Cleaning step after use (reference)

 
1. Disassemble the extraction vessele and take out the contents.
  
  HINT: If the contents do not come out, push it out with a stick.
 
2. Clean each part. (Extraction vessel, filter, body)

  Cleaning example:

    1) Rinse the residue with acetone.

    2) Soak in a cleaning solution and ultrasonically clean. (Water → Methanol → Acetone) 

  

  HINT: Since the extraction vessel is hard and heavy, a metal container is recommended as a cleaning container.

  HINT: Collect only the vessel with a metal colander.

  NOTE: If solid matter remains, it can be washed with a brush.
    Ultrasonic cleaning is desirable as it can cause leaks if the container is damaged. 

 

3. Dry

  

  NOTE: If the solvent remains, it will be mixed during extraction, so please dry it thoroughly.