(LC) Cleaning the extraction vessel (SFE)
(LC) Cleaning the extraction vessel (SFE)
- Category :
Answer
Cleaning step after use (reference)
1. Disassemble the extraction vessele and take out the contents.

HINT: If the contents do not come out, push it out with a stick.
2. Clean each part. (Extraction vessel, filter, body)
Cleaning example:
1) Rinse the residue with acetone.
2) Soak in a cleaning solution and ultrasonically clean. (Water → Methanol → Acetone)

HINT: Since the extraction vessel is hard and heavy, a metal container is recommended as a cleaning container.
HINT: Collect only the vessel with a metal colander.
NOTE: If solid matter remains, it can be washed with a brush.
Ultrasonic cleaning is desirable as it can cause leaks if the container is damaged.
3. Dry

NOTE: If the solvent remains, it will be mixed during extraction, so please dry it thoroughly.