FAQ - Frequently Asked Questions
- Products Search > Continuous Monitoring Analysis > (GAS) The indicator drifts (CGT-7100)
(GAS) The indicator drifts (CGT-7100)
(GAS) The indicator drifts (CGT-7100)
- Category :
Answer
• There is a large change in ambient temperature. → Select a location with as little possible change in temperature.
• The inner surface of the cell is dirty. → Clean and purge the measurement cell. Contact a Shimadzu sales or service representative.
• Problem in sampling unit. → Perform checks and repairs on each part.
• Gas is leaking inside the analyzer. → Check the leak in the piping. Click here for "Gas leak checking (CGT-7100)".
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.