(PDA) Temperature of the spectrometer is unstable, or does not reach the set temperature.

(PDA) Temperature of the spectrometer is unstable, or does not reach the set temperature.
Category : 

Answer

« Cause »
  • Abnormal temperature of the spectrometer
  • The fan supplying hot air has a problem.
 
« Measure »
  PDA-8000: Confirm that the temperature of the analysis chamber is 15 °C to 30 °C. After that, shut down the system, and then restart it.
  • Click here for Long term shutdown
 
  • Click here for Startup after a long period shut down
  PDA-7000: Check that the room temperature is within 10 - 28°C and reset the AIRCON switch on the switch panel.
 