FAQ - Frequently Asked Questions
- Products Search > Elemental Analysis > Optical Emission Spectrometer (PDA) > (PDA) Temperature of the spectrometer is unstable, or doe...
(PDA) Temperature of the spectrometer is unstable, or does not reach the set temperature.
(PDA) Temperature of the spectrometer is unstable, or does not reach the set temperature.
- Category :
Answer
« Cause »
• Abnormal temperature of the spectrometer
• The fan supplying hot air has a problem.
• The fan supplying hot air has a problem.
« Measure »
PDA-8000: Confirm that the temperature of the analysis chamber is 15 °C to 30 °C. After that, shut down the system, and then restart it.
• Click here for Long term shutdown
• Click here for Startup after a long period shut down
PDA-7000: Check that the room temperature is within 10 - 28°C and reset the AIRCON switch on the switch panel.
If not resolved, contact a Shimadzu sales or service representative.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.