• Font Size
  • S
  • M
  • L
  • No : 6536
  • Date of Issue : 2021/07/06 10:54
  • Print

(PDA) Temperature of the spectrometer is unstable, or does not reach the set temperature.

(PDA) Temperature of the spectrometer is unstable, or does not reach the set temperature.
Category : 

Answer

« Cause »
  • Abnormal temperature of the spectrometer
  • The fan supplying hot air has a problem.
 
« Measure »
  PDA-8000: Confirm that the temperature of the analysis chamber is 15 °C to 30 °C. After that, shut down the system, and then restart it.
  • Click here for Long term shutdown
 
  • Click here for Startup after a long period shut down
  PDA-7000: Check that the room temperature is within 10 - 28°C and reset the AIRCON switch on the switch panel.
 

May I have your opinion?

Please provide your feedback below. If a reply is required, please use one of the four red inquiry buttons at the bottom of the screen.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.