(PDA) The spectrometer inner pressure has problem

PDA-8000: The spectrometer inner pressure is not within the green range.
PDA-7000: The Vacuum does not become OK.
Category : 

Answer

« Cause »
  • The O-ring of the condenser lens is not correctly installed.
  • The oil level in the vacuum pump is not the specified one.
  • The pump oil in the vacuum pump is soiled, lowering the exhaust ability.
 
« Measure »
  Perform maintenance and inspection of the condenser lens and vacuum pump.
  • Click here for Replacing the Pump Oil
 