- Products Search > Elemental Analysis > Optical Emission Spectrometer (PDA) > (PDA) The spectrometer inner pressure has problem
(PDA) The spectrometer inner pressure has problem
PDA-8000: The spectrometer inner pressure is not within the green range.
PDA-7000: The Vacuum does not become OK.
- Category :
Answer
« Cause »
• The O-ring of the condenser lens is not correctly installed.
• The oil level in the vacuum pump is not the specified one.
• The pump oil in the vacuum pump is soiled, lowering the exhaust ability.
• The oil level in the vacuum pump is not the specified one.
• The pump oil in the vacuum pump is soiled, lowering the exhaust ability.
« Measure »
Perform maintenance and inspection of the condenser lens and vacuum pump.
• Click here for Replacing the Pump Oil
If not resolved, contact a Shimadzu sales or service representative.