AOC-20s vial is not set[Error Code:6022]

There was no sample vial at the specified vial number position.
The following causes are possible.
• The specified vial No. was a position on the tray with no vial.
• The sample vials on the turret have been removed during analysis.
• Teaching position setting and the actual placement do not match (if used with a GC other than GC-2030).
Category : 

Answer

Check and perform the following.
• Check the vial No. setting and the actual vial placement.
• Check the teaching position setting (if used with a GC other than GC-2030).