AOC-20s sample vial removed[Error Code:6021]
The following causes are possible.
• Grip was not firm enough. Vial dropped when it was being lifted.
• AOC is set to use the two-dimensional code reader while there is no two-dimensional code reader attached.
• Grip was not firm enough. Vial dropped when it was being lifted.
• AOC is set to use the two-dimensional code reader while there is no two-dimensional code reader attached.
- Category :
Answer
Check and perform the following.
• Redo the teaching using the vial to be used for actual analysis.
• Redo the teaching using the vial to be used for actual analysis.
[In case of AOC-30i]
AOC-30i_AOC-20sU_SamplerTeaching.pdf
AOC-30i_AOC-20sU_SamplerTeaching.pdf
[In case of AOC-20i]
AOC-20i_AOC-20s_SamplerTeaching.pdf
AOC-20i_AOC-20s_SamplerTeaching.pdf
• Check the two-dimensional code reader configuration.