- Products Search > GC > GC-2030 > Error/Warning > AOCm vial sensor error[Error Code:6040 6045]
AOCm vial sensor error[Error Code:6040 6045]
The following causes are possible.
• Power was turned on while a vial is exposed to direct sunlight.
• There is adhesion of dust or particles on the vial sensor.
• The vial sensor has a problem.
• Power was turned on while a vial is exposed to direct sunlight.
• There is adhesion of dust or particles on the vial sensor.
• The vial sensor has a problem.
- Category :
Answer
Check and perform the following.
• Check whether the operating environment is appropriate.
• Check whether the operating environment is appropriate.
• Remove any dust or particles from the vial sensor.