FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > AOCm can not start[Error Code:6004 6026]
AOCm can not start[Error Code:6004 6026]
Start is attempted under the following conditions:
• Door is open.
• Error is present.
• Power was turned on with the door open, and the [RESET] button is not pushed.
• The syringe replacement procedure has not been completed.
• The turret type and the GC sampler presence setting are not consistent.
• Door is open.
• Error is present.
• Power was turned on with the door open, and the [RESET] button is not pushed.
• The syringe replacement procedure has not been completed.
• The turret type and the GC sampler presence setting are not consistent.
- Category :
Answer
Close the door, remove the cause, and press the [RESET] button.
If the error occurs again after taking the measures, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
If the error occurs again after taking the measures, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.