• Font Size
  • S
  • M
  • L
  • No : 6048
  • Date of Issue : 2021/05/18 15:33
  • Last Updated : 2021/06/11 14:15
  • Print

AOCm plunger error[Error Code:6003 6025]

The following causes are possible.
• Plunger is bent.
• Plunger is stuck.
• The plunger motor or sensor has a problem.
Category : 

Answer

• If the plunger is bent, replace the syringe.
• If the plunger is not working properly, clean the plunger or replace the syringe.
• Revise the wash count and/or wash method as necessary.
   - If the sample remains inside the syringe: Increase the number of washes. Increase the types of solvents used for cleaning.
   - If the plunger is worn: Reduce the number of cleanings.
• If no abnormality is found in the plunger, and/or if the error occurs again after taking the measures, turn off the main power of the device and contact our sales office, distributor, or service representative designated by us.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.