• Font Size
  • S
  • M
  • L
  • No : 6034
  • Date of Issue : 2021/05/18 15:33
  • Last Updated : 2021/06/03 12:59
  • Print

DCG flow rate is not enough[Error Code:4213]

The pressure or flow rate may have not reached the set value due to the following reasons.
  • The gas supply pressure is insufficient.
  • There is a leak at tubing connections.
  • The resistance of supply tubing is large (since the tubing connected is long or its inner diameter is small).
Category : 

Answer

  • Check that the gas main valve is open.
  • Check that the gas supply pressure is within the range below.
  • Check that the supply pressure is sufficient for the BID detector gas to flow.
  • When using the helium purifier HP2, check that there is no leak at tubing connections.
  • Check that no tubing connections are loose and there is no leak.
  • Contact your Shimadzu service representative to check the inner diameter and length of used tubing.
 

May I have your opinion?

Please provide your feedback below. If a reply is required, please use one of the four red inquiry buttons at the bottom of the screen.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.