- Products Search > GC > GC-2030 > Error/Warning > Detected hydrogen leak[Error Code:51]
Detected hydrogen leak[Error Code:51]
Hydrogen gas leaks.
- Category :
Answer
- Stop the hydrogen gas supply.
- Check that the column connections have no leak.
- Check that the column is not broken.
- Check the tubing connections when using hydrogen gas in a system using AUX-APC, such as Advanced Flow Technology.
If the error appears even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.