Detected hydrogen leak[Error Code:51]

Hydrogen gas leaks.
Category : 

Answer

  • Stop the hydrogen gas supply.
  • Check that the column connections have no leak.
  • Check that the column is not broken.
  • Check the tubing connections when using hydrogen gas in a system using AUX-APC, such as Advanced Flow Technology.
If the error appears even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.