FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > DetAPCn ECDG was overpressure[Error Code:2268 2269 2270 2...
DetAPCn ECDG was overpressure[Error Code:2268 2269 2270 2271]
ECD detector gas flow exceeded the specified value. These errors may occur when supply gas pressure is changed during APC operation.
- Category :
Answer
- Do not open or close the main valve of the supply gas during APC operation.
- Restart the GC with the gas supplied.
- Check that the ECD gas supply pressure is within the range below.
- Check that the gas supply pressure is stable.
If the error appears repeatedly even after checking the above, the control system of the APC may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.