FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > DAMCn makeup set point not reached[Error Code:2272 2273 2...
DAMCn makeup set point not reached[Error Code:2272 2273 2274 2275]
The pressure or flow rate may have not reached the set value due to the following reasons.
- The gas supply pressure is insufficient.
- There is a leak at tubing connections.
- The resistance of supply tubing is large (since the tubing connected is long or its inner diameter is small).
- Category :
Answer
- Check that the gas main valve is open.
- Check that the gas supply pressure is within the range below.
- Check that the supply pressure is sufficient for the makeup gas to flow.
- Check that no tubing connections are loose and there is no leak.
- Contact your Shimadzu service representative to check the inner diameter and length of used tubing.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.