FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > CARn was overpressure[Error Code:2214 2215 2216 2217]
CARn was overpressure[Error Code:2214 2215 2216 2217]
Excessively high pressure is applied to the injection port. These errors may occur when supply gas pressure is changed during AFC operation.
- Category :
Answer
- Do not open or close the main valve of the supply gas during AFC operation.
- Restart the GC with the gas supplied.
- Check that the gas supply pressure is within the range below.
- Check that the gas supply pressure is stable.
If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.

May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.