FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > Analysis waiting[Error Code:5112]
Analysis waiting[Error Code:5112]
The analysis was started while the PREP RUN function is active and the instrument is in the waiting status.
- Category :
Answer
Set [Prep Run Wait Time] longer.
The PREP RUN is activated when using the following functions.
The PREP RUN is activated when using the following functions.
- Splitless (split in standby)
- Carrier gas saver (saved in standby)
- CRG (coolant blowing stops in standby)
- High pressure injection (normal pressure in standby)
- Back flush (back flush in standby)
The instrument will go into a standby state if no analysis starts when the [Prep Run Wait Time] elapses after the analysis preparation starts, due to extended pre-run time, etc. It is therefore necessary to set [Prep Run Wait Time] longer than pre-run time. For headspace analysis, set [Prep Run Wait Time] longer than heat-retention time.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.