FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > DET#n FTD overcurrent detected[Error Code:4105 4106 4107 ...
DET#n FTD overcurrent detected[Error Code:4105 4106 4107 4108]
Detector current is abnormal. If the resistance of the filament becomes abnormally high when the FTD is in use, the protection circuit is actuated to prevent damage to the filament.
- Category :
Answer
Turn off the main power of the instrument, and confirm the following items.
- The set value of the current exceeds the maximum operating current.
- The makeup gas is not flowing.
- A gas leak occurs.
Remove the cause of the error, then turn on the instrument. If the error cannot be resolved, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.