FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > APCy valve not completely closed[Error Code:2069 2070 207...
APCy valve not completely closed[Error Code:2069 2070 2071 2072 2073 2074 2075 2076 2077 2078 2079 2080 2081 2082 2083 2084 2085 2086]
There is a leak inside the pressure control valve of the AUX-APC.
- Category :
Answer
・Check that the tubing connection destination of the AUX-APC is not blinded. If it is blinded, remove the blind to open the line to atmosphere.
・If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
・If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.