DetAPCn DCG set point not reached[Error Code:2144 2145 2146 2147]
The pressure or flow rate may have not reached the set value due to the following reasons.
・No column is connected.
・The gas supply pressure is insufficient.
・There is a leak at tubing connections.
・The resistance of supply tubing is large (since the tubing connected is long or its inner diameter is small).
・No column is connected.
・The gas supply pressure is insufficient.
・There is a leak at tubing connections.
・The resistance of supply tubing is large (since the tubing connected is long or its inner diameter is small).
- Category :
Answer
・Check that the column is connected.
・Check that the gas main valve is open.
・Check that the gas supply pressure is within the range below.
・Check that the gas main valve is open.
・Check that the gas supply pressure is within the range below.
・Check that the supply pressure is sufficient for the detector gas to flow.
・Check that no tubing connections are loose and there is no leak.
・Contact your Shimadzu service representative to check the inner diameter and length of used tubing.
・Contact your Shimadzu service representative to check the inner diameter and length of used tubing.