• Font Size
  • S
  • M
  • L
  • No : 5974
  • Date of Issue : 2021/05/18 15:35
  • Last Updated : 2021/06/04 13:37
  • Print

SCD ozone source oxygen control irregularity[Error Code:6175]

The flow of the flow controller to control oxygen gas for the SCD ozone generator is unstable.
Category : 

Answer

・Check that the oxygen gas supply pressure is within the range below.
 
 
・Check that the gas supply pressure is stable.
・If the error appears repeatedly even after checking the above, the flow controller may have failed.
   Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Technology Inc.