- Products Search > GC > GC-2030 > Error/Warning > DET#n set temperature is low[Error Code:1049 1050 1051 1052]
DET#n set temperature is low[Error Code:1049 1050 1051 1052]
Detector temperature is set at a lower value than column temperature.
- Category :
Answer
・To prevent detector contamination, set a detector temperature higher than the column oven temperature.
・When setting a detector temperature lower than the column oven temperature, turn off the contamination prevention function.
・When setting a detector temperature lower than the column oven temperature, turn off the contamination prevention function.
・[HOME] - [Function] - [Configuration] - [Max Temperature] to display [Max Temperature] screen.
・Display the second page and turn off the contamination prevention function of the relevant detector.
・Display the second page and turn off the contamination prevention function of the relevant detector.