- Products Search > GC > GC-2030 > Error/Warning > COL CRG is out of control[Error Code:1034]
COL CRG is out of control[Error Code:1034]
Oven door opens.
The instrument cannot cool column oven because coolant becomes empty.
The instrument cannot cool column oven because coolant becomes empty.
- Category :
Answer
・When oven door opens, close the door and press [Reset].
・When coolant becomes empty, turn off the main power of the instrument and replace coolant.
・If the error appears even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
・When coolant becomes empty, turn off the main power of the instrument and replace coolant.
・If the error appears even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.