FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > Heat is escaping[Error Code:1001]
Heat is escaping[Error Code:1001]
Oven door opens.
Heat loss is large because the insulation of the oven has a hole.
- Category :
Answer
・When oven door opens, close the door and press [Reset].
・When the insulation of the oven has a hole, turn off the main power of the instrument and close the hole.
・If the error appears even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
・When the insulation of the oven has a hole, turn off the main power of the instrument and close the hole.
・If the error appears even after checking the above, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.