• Font Size
  • S
  • M
  • L
  • No : 5956
  • Date of Issue : 2021/05/18 15:37
  • Last Updated : 2021/06/03 17:16
  • Print

SCD exhaust for vacuum error[Error Code:6113]

A leak has occurred inside the SCD.
Category : 

Answer

Determine the cause of the leak by referring to the table below. 
If the error persists, turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
 
Cause Solution
The inner pyrotube is broken.
Replace the inner pyrotube.
The back ring is not installed on the inner pyrotube.
Install the back ring on the inner pyrotube.
The ferrule of the inner pyrotube is not inserted correctly.
Check that the ferrule is inserted in the inner pyrotube and that there is no abnormality in the ferrule.
The column is not properly tightened.
・Check the tightening condition of the nut.
・Additionally tighten it.
The vacuum pump is not turned on. Turn on the switch.
The degree of vacuum is inadequate.
Maintain the vacuum pump. If the problem is not resolved after maintaining the vacuum pump, replace the vacuum pump.
There is a leak between the vacuum pump and ozone scrubber.
Additionally tighten the connection.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.