PURn is not controlled [Error Code:9055 9056 9057 9058]

The analysis was started when the purge control was set to [Off].
Category : 

Answer

Set [Septum Purge] to [On] on [Purge Flow Program] screen, and then start the analysis.
 
<GC instrument monitor setting>
 
Press [HOME]  > [Injection port] > [Submenu] > [Purge Flow Program],
and then display [Purge Flow Program] screen and turn [On] septum purge. 
 
<Labsolutions setting> 
Set the purge flow rate control in the device monitor window to [On].
 