Can not switch gas because of gas selector# sub error occurring [Error Code:2372 2373]

You tried to set the supply gas to sub while there is an error with the sub of the gas selector.
Category : 

Answer

Check whether the sub gas is supplied and whether gas is not leaking from connections.
After resolving the problem, press [Reset].
 
If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply,
and contact your Shimadzu sales/service representative.