Can not switch gas because of gas selector# main error occurring [Error Code:2370 2371]

You tried to set the supply gas to main while there is an error with the main of the gas selector.
 
Category : 

Answer

Check whether the main gas is supplied and whether gas is not leaking from connections.
After resolving the problem, press [Reset].
 
If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply,
and contact your Shimadzu sales/service representative.