• Font Size
  • S
  • M
  • L
  • No : 5941
  • Date of Issue : 2021/05/18 15:36
  • Last Updated : 2021/06/09 14:38
  • Print

Can not switch gas because of gas selector# main error occurring [Error Code:2370 2371]

You tried to set the supply gas to main while there is an error with the main of the gas selector.
 
Category : 

Answer

Check whether the main gas is supplied and whether gas is not leaking from connections.
After resolving the problem, press [Reset].
 
If the error appears repeatedly even after checking the above, turn off the main power of the instrument, stop gas supply,
and contact your Shimadzu sales/service representative.

May I have your opinion?

Please provide your feedback below. This is for feedback only. You will not receive a response.
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.