FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > CARx septum counter warning [Error Code:9004 9005 9006 9052]
CARx septum counter warning [Error Code:9004 9005 9006 9052]
Exceeded the set threshold of inlet septum counter.
- Category :
Answer
Replace the septum and then reset the counter.
Septum replacement
Counter reset
<GC instrument monitor setting>
Press HOME button > [Function] > [Diagnosis] > [Analysis counter],
and then display [Analysis counter] screen and reset the liner usage counter.
![](/usr/file/attachment/AnalysisCounter.png)
<Labsolutions setting>
Press [System Check]> [Consumables Reset] on the assistant bar, and reset the counter from the [Reset] button.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.