CARm is not controlled [Error Code:2141 2142 2143 2213]

Temperature control started when the carrier gas control was turned off.
Category : 

Answer

Turn on the carrier gas control or remove unused carrier gas line from the analytical line.
 
How to set Flow control [On/OFF]
Press (HOME) - [Injection Port], and display bellow screen.
Set turn on bellow ❶. 
When use several analytical line, change analytical line.
 
How to change analytical line
(HOME)-[Settings]-[Submenu]-[LINE Configuration] to display [LINE Configuration] screen.
Remove check for unused units.