CARx calc. flow out of range [Error Code:5009 5010 5011 5073 5074 5077 5078 5102 5106 5107 5110 5111]
The total flow calculated from the inputted split ratio, column inlet pressure, column flow, linear velocity, and purge flow is outside the set range.
- Category :
Answer
Set the split ratio, column flow, and purge flow so that the total flow is within the AFC control range.
Total flow is calculated with the following equation:
(Total flow) = (Column flow) + (Column flow * Split ratio) + (Purge flow)
(Total flow) = (Column flow) + (Column flow * Split ratio) + (Purge flow)