- Products Search > GC > GC-2030 > Error/Warning > TCD signal was out of range [Error Code:4201]
TCD signal was out of range [Error Code:4201]
Zero adjustment cannot be performed because the TCD baseline is unstable.
The detector controller unit may have failed.
The detector controller unit may have failed.
- Category :
Answer
Perform zero adjustment again after the TCD baseline becomes stable.
If it does not work, press (HOME)-[Detector] and change to the optimum current value from the screen below.
* If using a packed TCD, please refer to the maintenance help.
When the problem persists after troubleshooting,Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.