FAQ - Frequently Asked Questions
- Products Search > GC > GC-2030 > Error/Warning > AOCn syringe counter error [Error Code:88 89]
AOCn syringe counter error [Error Code:88 89]
The number of analysis exceeded the set threshold.
- Category :
Answer
Replace the consumables and then reset the counter on [Analysis Counter] screen.
*Refer to the following for how to reset the counter value.
<When executing from the GC main unit screen>
Press (HOME) - [Function] - [Diagnosis] - [Analysis Counter] to display [Analysis Counter] screen.
Press [Reset] to reset the count.
Press (HOME) - [Function] - [Diagnosis] - [Analysis Counter] to display [Analysis Counter] screen.
Press [Reset] to reset the count.

<When executing from the LabSolutions>
Press [System Check]> [Reset Consumables] on the assistant bar, and use the Syringe Usage Counter [Reset] button to reset.
Press [System Check]> [Reset Consumables] on the assistant bar, and use the Syringe Usage Counter [Reset] button to reset.
May I have your opinion?
This site uses the FAQ system "PKSHA FAQ", which is a product of PKSHA Communication Inc.